Accuracy of an Automatic Infusion Controller (AutoClamp) for Intravenous Fluid Administration
Geun Joo Choi, Il Jae Yoon, Oh Haeng Lee, Hyun Kang*
Article Information
Identifiers and Pagination:
Year: 2015Volume: 9
First Page: 23
Last Page: 28
Publisher ID: TOATJ-9-23
DOI: 10.2174/1874321801509010023
Article History:
Received Date: 15/12/2014Revision Received Date: 09/08/2015
Acceptance Date: 19/08/2015
Electronic publication date: 7/10/2015
Collection year: 2015
open-access license: This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International Public License (CC-BY 4.0), a copy of which is available at: https://creativecommons.org/licenses/by/4.0/legalcode. This license permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Abstract
Background:
This study was purposed to compare in vitro the volumetric accuracy of a newly introduced automatic infusion controller, AutoClamp with that of other commonly used infusion devices.
Methods:
Four different volumetric infusion devices were used to examine the accuracy: Terufusion TE-112; Volumed µVP7000; AutoClamp; and Infucon. Accuracy was determined for each flow rate (20, 40, 100, and 200 ml/h) by using infusate volumes collected after 3 h of initiating the fluid administration. Accuracy was calculated as the percentage difference between set volume and actual volume delivered. The influences of fluid viscosity and flow resistance on infusion device accuracy were also evaluated.
Results:
There were no cases of a greater-than-10% difference between set volume and actual delivered volume. The accuracy of the Infucon was significantly less than that of the other devices. Infusion devices proved to be consistent and unaffected by fluid viscosity or flow resistance except for the Infucon.
Conclusion:
The accuracy of the AutoClamp was comparable to that of other commonly used infusion pumps (Terufusion TE-112 and Volumed µVP7000) regardless of infusate viscosity and flow resistance.